INSTALLATION

14

OPTISENS PH 8390

www.krohne.com

03/2012 - 4001930601 MA OPTISENS PH 8390 R01 en

3.5.1 Connecting the sensor cable to the signal converter

When ordering the single channel version, only the interface "Pos.A" is populated. In the version
with dual channel the interfaces "Pos.A" and "Pos.B" are populated.

DANGER!
All work on the electrical connections may only be carried out with the power disconnected. Take
note of the voltage data on the nameplate!

INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.

Figure 3-3: Sensor connection terminals on the signal converter dual channel version with terminal block A+B

1 Sensor connection terminals
2 Terminal block S (protective earth) and, to the right, terminal block A
3 Terminal block A: terminals for sensor and temperature
4 Terminal block B: terminals for sensor and temperature

Wire

Wire

Wire

Wire

Terminal block Pos.A/B

Terminal block Pos.A/B

Terminal block Pos.A/B

Terminal block Pos.A/B

OPTISENS PH 8390 with 4-wire double coax cable

OPTISENS PH 8390 with 4-wire double coax cable

OPTISENS PH 8390 with 4-wire double coax cable

OPTISENS PH 8390 with 4-wire double coax cable

Inner coax shield (black)

N (ref.)

Coax core (transparent)

O (pH)

Pt100 (white)

P

Pt100 (red)

X