3.5.2 Cable assign of a Hirschmann plug
3.5.3 Connecting the sensor cable to the signal converter
CAUTION!
Do not shorten the cable length. This can influence the measuring accuracy.
INFORMATION!
The given colours only refer to the delivered cables! When using cables of a different provider,
take into account the manufacturer’s notes concerning the cable colour and designation.
Figure 3-2: Cable assign of a Hirschmann plug
1 Cable LIYCY 4x0.5 mm
2 White
3 Brown
4 Yellow
5 Green
6 Shield
7 Pt100 sensor
8 Outer electrode
9 Inner electrode
DANGER!
All work on the electrical connections may only be carried out with the power disconnected. Take
note of the voltage data on the nameplate!
CAUTION!
Don’t connect the shielding to the signal converter if the sensor has got a Hirschmann GDM 3011
plug!
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.