KROHNE UFM 500 EN User Manual

Page 36

Advertising
background image

36

continued

yes

yes

Check direction of flow, see sect.

5.4. Do all receiver instruments

deflect in the right direction?

Do receiver instruments indicate

the right value?

Is local display steady at

constant flowrate?

yes

no

no

yes

System in order.

yes

Connect receiver instruments with

correct polarity!

no

UFC 400… signal converter:

Consult factory!

UFC 500… signal converter:

Increase time constant, fct. 3.3.6.

Reading steady?

no

yes

Check the outputs as described in

sect. 7.1.2 + 7.1.3. In order ?

no

Replace electronic unit or

consult factory.

UFC 500… signal converter:

Test display, fct. 2.1.

In order?

no

yes

yes

Advertising