KROHNE BATCHCONTROL 5014 C User Manual
Page 24

Installation and operating instructions
24 BATCHCONTROL
6.4 Error limits at reference conditions
F
= Error in % of MV
MV
= measured value
Pulse output
DN 2.5 – 6
/
1
/
10
“ –
1
/
4
“
DN 10 – 40
/
3
/
8
“ – 1
1
/
2
“
at flow velocity of ...
v
≥ 1 m/s ≥ 3.3 ft/s
F < ± 0.5 % of MV
F < ± 0.3 % of MV
v < 1 m/s < 3.3 ft/s
F < ± 0.4 % of MV + 1 mm/s
F < ± 0.2 % of MV + 1 mm/s
< ± 0.4 % of MV + 0.04 inch/s
< ± 0.2 % of MV + 0.04 inch/s
Repeatability Filling time T
F
Standard
deviation
σ
1.5 s < T
F
≤ 3 s
≤ 0.4 %
3.0 s < T
F
≤ 5 s
≤ 0.2 %
5.0 s < T
F
≤ 0.1 %
Reference conditions (similar to EN 29 104)
Liquid product
water +20 °C / +68 °F
Straight inlet/outlet runs
10× DN / 5 × DN (DN = meter size)
Valve closing time variation
< 1 ms
Flow velocity
1 m/s = 3.3 ft/s
Volumetrically wet calibration on EN 17025 accredited calibration rigs.
6