6
TECHNICAL DATA
24
OPTIFLUX 5000
www.krohne.com
04/2011 - 4000686302 - HB OPTIFLUX 5000 FL R02 en
Operating conditions
Temperature
Temperature
Temperature
Temperature
Process temperature
Compact version: -40...+140°C / -40...+284°F
Remote version: - 40...+180°C / -40...+356°F
For Ex versions different temperatures are valid.
Please check the relevant Ex documentation for details.
Maximum temperature
change (shock)
120°C / 248°F
Ambient temperature
-40…+65°C / -40…+149°F
For Ex versions different temperatures are valid.
Please check the relevant Ex documentation for details.
Storage temperature
-50…+70°C / -58…+158°F
Pressure
Pressure
Pressure
Pressure
Ambient
Atmospheric
Nominal flange pressure
Standard:
Standard:
Standard:
Standard:
EN 1092-1
DN200...300: PN10
DN100...150: PN16
DN15...80: PN40
ASME B16.5
Standard:
Standard:
Standard:
Standard:
1...12": 150 lb
½
": 300 lb
Option:
Option:
Option:
Option:
1...3": 300 lb
Vacuum load
0 mbar / 0 psi
Pressure ranges for
secondary containment
Pressure resistant up to 40 bar / 580 psi
Burst pressure up to approx. 160 bar / 2320 psi
Chemical properties
Chemical properties
Chemical properties
Chemical properties
Physical condition
Conductive liquids
Electrical conductivity
Non water:
Non water:
Non water:
Non water:
DN25...300: ≥ 1 μS/cm
DN15: ≥ 5 μS/cm
Demineralised cold water:
Demineralised cold water:
Demineralised cold water:
Demineralised cold water:
DN15...300: ≥ 20 μS/cm
Permissible gas content
(volume)
≤ 5%
Permissible solid content
(volume)
IFC 100: ≤ 10%
IFC 300: ≤ 70%
Recommended flow
velocity
-12...+12 m/s / -40...+40 ft/s
Installation conditions
Installation
Take care that the flow sensor is always fully filled.
For detailed information refer to
Installation
on page 10.
Flow direction
Forward and reverse.
Arrow on flow sensor indicates positive flow direction.