Electrical connections, 7 protection category – KROHNE DK 32-DK 34-DK 37 EN User Manual

Page 21

Advertising
background image

ELECTRICAL CONNECTIONS

4

21

DK32 - DK34 - DK37

www.krohne.com

06/2014 - 4000287103 MA DK32-34-37-R05-en

4.7 Protection category

The variable area flowmeters DK32 / DK34 / DK37 have been tested for ingress protection
according to EN 60529. The following grades of protection are confirmed:

Therefore it is essential to observe the following points:

• Use only original gaskets. They must be clean and free of any damage. Defective gaskets have

to be replaced.

• The electrical cables used must be undamaged and must comply with regulations.
• The cable has to be laid with a loop 3 upstream of the flowmeter to prevent water from

getting into the housing.

• The cable glands 2 have to be firmly tightened.
• Close the unused cable glands using blanking plugs 1.
• Do not remove the specified outer sheath from the cable gland.

Ingress Protection
EN 60529

Mechanical version
without cable glands

Electrical version
with cable glands

DK32 / DK34

IP66 / IP68

IP65

DK37

IP66

IP66

DANGER!
After all servicing and maintenance work on the flowmeter, the specified protection category has
to be ensured again.

Figure 4-8: Cable gland

1 Close the unused cable glands using blanking plugs.
2 The cable glands have to be firmly tightened.
3 The cable has to be laid with a loop.

MA_DK32_34_37_R05_en_287103_PRT.book Page 21 Friday, June 6, 2014 11:42 AM