Technical data – KROHNE DK 32-DK 34-DK 37 EN User Manual
Page 30

7
TECHNICAL DATA
30
DK32 - DK34 - DK37
www.krohne.com
06/2014 - 4000287103 MA DK32-34-37-R05-en
Approvals
M8E HART
®
parameter configuration
Name of manufacturer (code)
KROHNE Messtechnik (69)
Name of model
M8E (230)
HART
®
protocol revision
5.1
Device revision
1
Physical layer
FSK
Device category
Transmitter
M8E process variable
M8E process variable flow
Values [%]
Signal output [mA]
Over range
+102.5 (±1%)
20.24...20.56
Device error detection
>106.25
≥21.00
Maximum
112.5
22
Multi-drop operation
-
4.5
Min. U
ext.
14.8 VDC
Standard
Indicator
Designation
ATEX
DK32 DK34 mechanical
II2GD IIC
II3GD IIC
DK32 DK34 electrical
II2G Ex ia IIC T6
DK37 mechanical
II2GD IIC
II3GD IIC
DK37 electrical
II2G Ex ia IIC T6
IEC Ex
DK32 DK34 electrical
Ex ia IIC T6
FM
DK32 DK34
IS/I/1/ABCD;T6
NI/I/2/ABCD;T6
S/II, III/2/FG;T6
IS/I, II, III/1/A-G
NI/II/2/ABCD
Nepsi
DK32 DK34
Ex nA II T1-T6
DK37
Ex ia IIC T1-T6
MA_DK32_34_37_R05_en_287103_PRT.book Page 30 Friday, June 6, 2014 11:42 AM