Start-up
5.1 Start-up
CAUTION!
When starting up the flowmeter, the following points must be observed:
•
Compare the actual operating pressure and the process temperature of the system with the
specifications on the nameplate (PS and TS); these limits must not be exceeded.
•
Make sure materials are compatible.
•
Close the needle valve at the flowmeters.
•
Slowly open the shut-off valve upstream and downstream of the flowmeter.
•
When measuring liquids, vent the pipes carefully.
•
When measuring gases, increase pressure slowly.
•
void float impact (e.g. caused by solenoid valves), as this is likely to damage the measuring
section or float.
•
Open needle valve at the flowmeters and set the required flow rate.
•
The top edge of the float marks the reading line for flow values:
Figure 5-1: Reading edge
1 Reading edge
2 Floatform AIII
3 Floatform ball