Technical data – KROHNE OPTIMASS 2000 EN User Manual
Page 36

6
TECHNICAL DATA
36
OPTIMASS 2000
www.krohne.com
10/2010 - 4000810002- MA OPTIMASS 2000 R02 en
ATEX (acc. 94/9/EC) temperature limits
Maximum end loadings
Ambient temp.
T
amb
°C
Max. medium
temp. T
m
°C
Temp. class
Max. surface
temp. °C
OPTIMASS 2000 / 2010C with or without
heating jacket / insulation
40
65
T6
T80
75
T5
T95
110
T4
T130
130
T3-T1
T150
65
75
T5
T95
110
T4
T130
130
T3-T1
T150
OPTIMASS 2300C Aluminium converter
housing - with or without heating jacket
/ insulation
40
50
T6
T80
65
T5
T95
100
T4
T130
130
T3-T1
T160
50
65
T5
T95
100
T4-T1
T130
60
60
T4-T1
T90
65
1
65
T4-T1
T95
OPTIMASS 2300C Stainless Steel
converter housing - with or without
heating jacket / insulation
40
50
T6
T80
65
T5
T95
100
T4
T130
120
T3-T1
T150
50
65
T5
T95
75
T4-T1
T105
55
55
T5-T1
T85
1 depending on I/O option. Please call for more information.
S100
S150
S250
Flanges
Flanges
Flanges
Flanges
20°C
40 barg
150kN
350kN
550kN
100 barg
100kN
120kN
60kN
150 barg
180 barg
130°C
32 barg
150kN
280kN
400kN
80 barg
60kN
50kN
50kN
115 barg
130 barg
Hygienic (all connections)
Hygienic (all connections)
Hygienic (all connections)
Hygienic (all connections)
130°C
10 barg
5kN
-
-