KROHNE BM 700 EN User Manual

Page 15

Advertising
background image

BM 700 Installation and Operating Instructions Page 15

Sequence for setting parameters (example)
(for version with local display)

The following description refers to a storage tank with the parameter examples taken
from the illustration in this Section. If the device no longer contains the default
parameters, the keystroke combination for entering the numerical values may differ.

Activity

Keys to be

actuated

Content of BM 700 display

after activity carried out

Entry into configuration menu

→

Fct. 1.0

OPERATION

Setting the parameter: tank height

↑

↑

→

→

Fct. 3.1.1

Tankheight

Display of default value

→

10.000 m

Input of tank height "5.30 m"

→

9x

↑

→

5x

↑

→

3x

↑

05.300 m

Confirm tank height and move to block
distance

↵

↑

Fct. 3.1.2

Blockdist

Display default value

→

0.5000 m

Enter block distance "0.60 m"

→

↑

0.6000 m

Confirm block distance and move to current
output configuration

↵

↵

↑

↑

Fct. 3.3

SIGNAL OUT

Move to lower range value

→

↑

↑

Fct. 3.3.3

Scale 4 mA

Display default value

→

+ 00.000 m

Enter lower range value (0.4 m = 4 mA)

3x

→

4x

↑

+ 00.400 m

Confirm lower range value and move to full-
scale range value

↵

↑

Fct. 3.3.4

Scale. 20mA

Display of default value

→

010.00 m

Enter full-scale value (4.0 m = 20 mA)

2x

→

9x

↑

→

4x

↑

004.00 m

Confirm full-scale value and move to empty
tank spectrum

↵

↵

↑

↑

→

↑

Fct. 3.5.2

EMPTY.SPEC.

Select: re-record empty spectrum

→

↑

↑

RECORD

Store changed parameters

↵

Accept. Yes

Confirm and select: averaging

↵

↑

AveragE

Confirm and start recording; then wait for
approx. 2 minutes!

↵

↵

READY

Confirm and move to tank type

↵

7x

↑

Fct. 3.5.9

Tank type

Display of default value

→

PROC tank

Select tank type "storage tank"

↑

↑

Storage t.

Return to measurement function with
confirmation of changed parameters

5x

↵

Param.Check,

then START,

then meas.val. display