BW25
www.krohne.com
09/2010 - 4000432302 MA BW25 R05 en
6.2.3 Changing the settings in the menu
Starting operation
Starting operation
Starting operation
Starting operation
Operation is started using the → key
If a different key is pressed, it is necessary to wait 5 seconds before activating the → key.
If a control lockout is set, the code → → → ^ ^ ^ ↑ ↑ ↑ must be entered. If no key is pressed
within 5 seconds, code input is exited.
Exiting operator input
Exiting operator input
Exiting operator input
Exiting operator input
Operation is exited by pressing the ^ key several times.
Example: changing the default parameter from m
m
m
m to cm
cm
cm
cm
6.2.4 Measures in the event of faulty indications
If the indications on the display or the responses to keypad commands are faulty, you have to do
a hardware reset. Switch the power supply OFF and ON again.
If data have been changed:
Save Yes
→
Changes saved.
An update is carried out and the display returns to measuring mode.
Save No
^
^
^
^
Changes not saved.
The display returns to measuring mode.
CAUTION!
Each time parameters or settings are changed, the measuring device carries out an internal
plausibility check.
If implausible inputs have been made, the indicator remains in the current menu, and the
changes are not accepted.
Display
Display
Example:
5.0
m
1x →
Fct. 3.13.1111
LEVEL
1x →
Fct. 1111.0
OPERATION
1x →
5.0000
m
2x ↑
Fct. 3333.0
INSTALLATION
6x ↑
500
cm
1x →
Fct 3.1111
LANGUAGE
1x ^
^
^
^
Save Yes
12x ↑
Fct 3.13
13
13
13
END&UNIT
3x ^
^
^
^
500
cm