KROHNE ACM 500 EN User Manual
Page 10

ACM 500
02/2013
10
Adjustable measuring ranges conductivity:
No. Range
Resolution
1
0... 0.5 ms
0.001 ms
2
0... 1 ms
0.001 ms
3
0... 2 ms
0.010 ms
4
0... 3 ms
0.010 ms
5
0... 5 ms
0.010 ms
6
0... 10 ms
00.10 ms
7
0.. . 20 ms
00.10 ms
8
0... 30 ms
00.10 ms
9
0... 50 ms
00.10 ms
10
0...100 ms
000.1 ms
11
0...200 ms
000.1 ms
12
0...300 ms
000.1 ms
13
0...500 ms
000.1 ms
14
0...999 ms
000.1 ms
Ranges set in as-delivered condition:
Meas. range
mS
1 0...200
2
0...20
3
0...2
4
0...0.5
3.4
Setting of temperature compensation 2, 4, 6, 8
Each measuring range is assigned its own setting for temperature compensation. The range is
adjustable between 0%/K (no compensation) and max. 5%/K. The compensation calculator
operates linearly on the basis of a reference temperature of 25°C. When supplied, the setting for
all ranges is 2%/K.
1
mS
Setting temperature coefficient
20
40
60
80
Meas. range No.
0... %/K
Measured conductivity
1
mS
Setting for TC 0...5.0 %/K
(with Jog Shuttle)
21
41
61
81
Meas. range No.
%/K
^
v
Measured conductivity