ACM 500

02/2013

6


2

Electrical connection

Terminals 1(+) and 2(-) are used for supplying a DC voltage of 18...36V. Terminal 2 is connected
to the housing via a protective diode. The maximum power consumption is 180 mA. This value
should be taken into account for the recommended fuse. An active 4...20 mA current output,
galvanically isolated from the supply voltage, is available at each of terminals 3 and 4, and 5 and
6. Terminal pair 3/4 supplies the conductivity signal, terminal pair 5/6 the temperature value. The
negative terminals of current outputs 4 and 6 are connected to each other internally. 24-V control
signals (pnp) can be connected to terminals 7 (R2) and 8 (R1) for external selection of one of the
four adjustable measuring ranges. The ground reference is connected to terminal 2; an open
terminal signifies 0 V.

Meas. Range

R2

R1

1

0 V

0 V

2

0 V

24 V

3

24 V

0 V

4

24 V

24 V

Please note currently valid installation regulations.