4/10

Error messages

Description of error

Error clearance

LINE INT.

Power failure Note:

Cancel error in RESET-QUIT menu

no counting during power failure

Reset totalizer if necessary.

OVERFLOW I

Current output overranged.

Check and if necessary correct
instrument parameters. After
elimination of cause, error message
deleted automatically.

OVERFLOW P

Pulse output overranged.

Check and if necessary correct

Note:

instrument parameters. After

totalizer deviation possible.

elimination of cause, error message
deleted automatically.

TOTALIZER

Totalizer has been reset

Cancel error message in
RESET/QUIT. menu.

ADC

Analog / digital

Error message deleted automatically

converter overranged

after elimination of cause.

FATAL. ERROR

Fatal error, all outputs set

Please consult factory.

to “min. values”

EMPTY PIPE

Pipe has run dry.

Fill pipe.

This message appears only when
the “empty pipe identifier”
option is installed
and the function
is switched on under Fct. 3.06
APPLICAT., submenue
“EMPTY PIPE”.

Error messages in measuring mode

4.5

The following list gives all errors that can occur during process flow measurement.
Errors shown in display when “YES” set in Fct. 1.04 DISPLAY, subfunction “DISP. MSG.”: