SERVICE

7

27

OPTIFLUX 2070

www.krohne.com

04/2008 • 4000243601

•

Replace the cover.

•

Tighten the 4 bolts.

•

Enter the programming mode, hold the V and Z key for 5 seconds.

i

The display flashes.

•

Press the Z key to enter the menu.

•

Scroll through the available positions with the V key to position 74 and check the battery

capacity (important for battery life indication).

•

Go to position 75 and enter "1" (reset battery counter).

i

Battery capacity symbol at the display should be "full".

•

Hold the V key for 3 seconds to confirm the new value and go back to the measuring mode.

•

Replace the blue protection cap.

WARNING!
Make sure that the battery cable is not jammed by the cover.