2. Electrical

installation

2.1

Location and connecting cables


Location

Do not expose the compact flow meter to direct sunlight. Install a sunshade if necessary.

Connecting cables

To conform to protection category requirements, observe the following recommendations:
–

Fit blanking plug PG 16 and apply sealant to unused cable entries.

–

In case if the terminal area is Eed, please use approved blanking plug.

–

Do not kink cables directly at cable entries.

–

Provide water drip point (U bend in cable).

–

Do not connect rigid conduit to cable entries.

–

If cables are a tight fit, enlarge inside diameter of cable gland by removing the appropriate
ring(s) from the seal.


2.2

Electrical Connection on converter without intrinsically safe outputs

–

Note information given on the instrument dataplate (voltage, frequency)!


–

Electrical connection in conformity with IEC 364 or equivalent national standard.


–

Special regulations apply to installation in hazardous areas. Please refer to separate "Ex"
installation instructions.


– The PE protective ground conductor must be connected to the separate U-clamp

terminal in the terminal box of the signal converter.


–

Do not cross or loop the cables in the terminal box of the signal converter. Use separate
(PG or NPT) cable glands for power and output cables.


– Ensure

that

the

screw thread of the round cover on the terminal box is well greased at

all times.

NOTE: The grease used must be non-corrosive to aluminium; typically it must be resin- and

acid-free.


– Protect sealing ring from damage.


5 6 4 4.1 4.2 11 12



N L AC
Outputs - + DC
connection PE
see Sect. 2.2
Power and signal connections for MFC 085 K

Please ensure that the information about power given on the
dataplate corresponds to the locally available mains voltage.