MFC010 Interface Manual

41

7.3

Discrete Input (Binary) Status Registers

The Discrete I nput S tatus r egisters ar e us ed t o i ndicate t o t he c ontrol s ystem t he c urrent s tate of t he

meter. From these values the user can access more detailed information in the other register groups.

Many of these states can also be accessed using the “Read Exception Status” command (Command No.

7).

Register No.

Description

1001

Supervisor Lock Password State

- The current state of the Supervisor Password protection. For details on activation,

de-activation and operation of the Supervisor password refer to Section 8.5 on page

76 for further details.

Range :

0 = Password Protection is Inactive

1 = Password Protection is Active

1002

Service Lock Password State

- The current state of the Service Password protection. For details on activation, de-

activation and operation of the Service password refer to Section 8.5 on page 76 for

further details.

Range :

0 = Password Protection is Inactive

1 = Password Protection is Active

1003

Custody Transfer Lock Password State

- The current state of the Custody Transfer Lock Password protection. For details on

activation, de-activation and operation of the Custody Transfer password refer to

Section 8.9 on page 78 for further information.

Range :

0 = Password Protection is Inactive

1 = Password Protection is Active

1004

Parameter Changed – Awaiting “Save Changes to EEPROM”

– This state indicates when configuration changes have been made but not yet

saved to the non-Volatile memory (EEPROM). See Section 8.6 on page 77 for a

more detailed explanation of saving and restoring configuration settings.

Range :

0 = No new data to save

1 = Data has been written but not stored in EEPROM

1005

System Error Flag

– This state indicates when the system has detected one or more error conditions

that may affect the operation of the meter. The individual System Error flags are

accessed from Input Register No 7001. (See Section 9 for more details on Errors

and Warnings)

Range :

0 = No system error flags detected

1 = One or more system error conditions are present

1006

Process Warning Flag

– This state indicates when the system has detected one or more process conditions

that may affect the operation of the meter. The individual Process warning flags are

accessed from Input Register No 7005. ( See section 9 for more details on Errors

and Warnings).

Range :

0 = No process warning flags detected

1 = One or more process warning conditions are present