3. SPECIFICATIONS

●

Measuring Range and Accuracy (at 23

±

5°C, relative humidity 45 -75%)

○

Insulation Resistance Ranges:

Nominal Output Voltage

250V

500V

1000V

Measuring

0 ~19.99M

Ω

0 ~19.99M

Ω

0 ~19.99M

Ω

Ranges

0 ~ 199.9M

Ω

0 ~ 199.9M

Ω

0 ~ 199.9M

Ω

0 ~ 1999M

Ω

0 ~ 1999M

Ω

0 ~ 1999M

Ω

Open-Circuit

250V DC

500V DC

1000V DC

Voltage

+20%,-0%

+20%,-0%

+20%,-0%

Nominal Current

1mA DC min.

1mA DC min.

1mA DC min.

at 0.25M

Ω

at 0.5M

Ω

at 1M

Ω

Short-Circuit

Current

1.5mA approx

20M

Ω

±

1.5%rdg

±

5dgt

200M

Ω

Accuracy

2000 0~1G

Ω

±

3%rdg

±

3dgt

M

Ω

1G~2G

Ω

±

10%rdg

±

3dgt

○

Continuity Ranges:

Ranges

20

Ω

200

Ω

2000

Ω

Measuring Ranges

0 ~ 19.99

Ω

0 ~ 199.9

Ω

0 ~ 1999

Ω

Open-Circuit

Voltage

7 ~ 12V approx

Measuring Current

at 0.2

Ω

~ 2

Ω

200mA min

Accuracy

±

1.5%rdg

±

5dgt

±

1.5%rdg

±

3dgt

○

AC Voltage Indication

0 ~ 600V

±

5%rdg

±

3dgt

―

5

―

2. FEATURES

MODEL-3005A/3007A are microprocessor controlled insulation-continuity

testers.

●

Designed to safety standards:

IEC 61010-1 Overvoltage CAT III 300V Pollution Degree 2

IEC 61010-2-31 (Safety requirements for hand-held probe assemblies)

●

Display with back light function to facilitate work at night or dimly lit

locations (Model 3007A only)

●

Bar graph to indicate measured results

●

Strap belt to make both hands' operation easier

●

Live circuit warning indication and buzzer

●

Auto discharge function

When insulation resistance is measured, electric charges stored in

capacitive circuits are automatically discharged after testing. Discharge

can be checked with live circuit warning.

●

Auto power off function

To prevent the instrument from being left powered on and conserve

battery power, the instrument automatically turns off approx. 10 minutes

after the last switch operation.

●

LOK mode (Model 3007A only)

The test current is removed once a stable reading is reached to prevent

unnecessary battery consumption.

―

4

―