Martel Electronics BetaGauge 321A-Ex User Manual

Page 23

Advertising
background image

21

4. Use SELECT to toggle through the UNIT options, and select NEXT to move on.

5. Use the

↑ and ↓ arrows to set the 100% point of the desired pressure range, select

DONE SET when finished.

6. Again, use the arrows to set 0% point and select DONE SET when finished and the

%ERROR mode will be ready to use.