Mechanical

Dimensions 4.5” (diameter) x 2.2” (depth) x 5” (height)

(11.4cm x 5.6cm x 12.7cm) (PI PRO depth is 1.5” or 3.7cm)

Pressure Connection: ¼” NPT Male

Housing: Stainless steel, meets NEMA 4/IP65 (PI PRO: cast Z

N

AL)

Display

5-1/2 Digits, 0.65” (16.53 mm) high

20-Segment bar graph, 0 to 100%

Power

Battery three (3), size AA alkaline batteries

Battery Life 1,500 hours without backlight (continuous on)

2,000 hours at slow sample rate

Low Battery Indicator icon is displayed near the end of battery life

Appendix 1: BetaGauge PI Calibration Procedure

Overview

Calibration adjustment of the BetaGauge PI is performed electronically
via internal software with the case closed. There are no mechanical
adjustments; all calibration commands and adjustments are done via
the keypad, using the display to guide the user through the calibration
process. Note that absolute range gauges require a factory calibration
and must be returned to the factory or an authorized service center.

Eight calibration points are used in the adjustment program, working
from full scale to zero at pressures equaling 100%, 87.5%, 75%, 62.5%,
50%, 37.5%, 25%, 12.5%, and 0% of full scale plus vacuum.

Note: This is an ambient temperature calibration, and should be
performed at an ambient temperature of 23° C ± 3° C (72° F ± 5° F).
Calibration outside this temperature range will invalidate the
temperature compensation program in the BetaGauge PI.

Calibration Interval

You should check performance of the BetaGauge PI at the interval
required by your calibration program. We recommend adjustment
when measurement deviates by more than 75% of the specified
accuracy, or 0.04%

Test Equipment

Verification and calibration of the BetaGauge PI requires pressure
and/or vacuum standards able to produce and indicate pressures from

9

Advertising