Martel Electronics BetaGauge PI PRO User Manual

Page 14

Advertising
background image

12

Use the Pressure Standard to output 125.00 psi (25%). After the
output has stabilized, press the ENTER key to continue. As the unit
takes readings, the screen will show ———-. When the readings are
complete the screen should look as shown below.

Use the Pressure Standard to output 62.50 psi (12.5%). After the
output has stabilized, press the ENTER key to continue. As the unit
takes readings, the screen will show ———-. When the readings are
complete the screen should look as shown below.

Use the Pressure Standard to output 0.00 psi. After the output has
stabilized, press the ENTER key to continue. As the unit takes
readings, the screen will show ———-. When the readings are
complete the screen should look as shown below.

Note: Only some ranges use vacuum calibration. If your gauge is not
one, than this step will be automatically skipped and calibration will be
complete.

Use the Pressure Standard to output -12.00 psi. After the output has
stabilized, press the ENTER key to continue. As the unit takes
readings, the screen will show ———-. When the readings are
complete the unit will reset and power up in normal mode.

PSI

-12.00

CAL MODE

PSI

0.00

CAL MODE

PSI

62.50

CAL MODE

Advertising
This manual is related to the following products: