Betagauge pi serial interface instructions, List of commands – Martel Electronics BetaGauge PI PRO User Manual

Page 16

Advertising
background image

14

9. BetaGauge PI Serial Interface Instructions

Initiating Communication

The terminal communications can be setup using terminal
communication software on a PC. The terminal settings need to be set
as follows:

• Bits per second: 9600
• Data bits: 8
• Parity: None
• Stop bits: 1
• Flow control: None
• Local echo on

List of Commands

Command

Description

CAL_START

Puts the calibrator in calibration mode

*CLS

Clears the error queue.

FAULT?

Returns an error code from the error queue

*IDN?

Identification query. Returns the manufacturer, model number,
and firmware revision level of the Calibrator.

TARE

Tares the offset pressure of the reading on the calibrator

TARE?

Returns the current tare value

PRES_UNIT?

Returns the pressure unit for the upper display.

PRES_UNIT

Sets the pressure unit for the display

ZERO_MEAS

Zeros pressure of the calibrator

ZERO_MEAS?

Returns the current zero offset value

MINMAX_RST

Resets the minimum and maximum recorded values.

MIN?

Returns the minimum recorded value

MAX?

Returns the maximum recorded value

HC_OFF

Turns unit off

HC_DFLT

Sets auto off defaults

TEMP?

Returns temperature in the chosen units

HC_COMP_OFF

Turns temperature compensation off.

HC_COMP_ON

Turns temperature compensation on.

HC_COMP?

Returns state of temperature compensation.

HC_RD_2410?

Return 2410 ADC counts.

HC_SI_OFF

Turns SI mode off.

HC_SI_ON

Turns SI mode on.

Advertising