Martel Electronics BetaProbe TI/TI+ User Manual

Page 3

Advertising
background image

This unit was designed to meet the above requirements using:

EN 60079-0:2006 ELECTRICAL APPARATUS FOR
EXPLOSIVE GAS ATMOSPHERES -- PART
0: GENERAL REQUIREMENTS (IEC
60079-0:2004 (MOD))

EN 60079-11:2007
EXPLOSIVE ATMOSPHERES -- PART 11:
EQUIPMENT PROTECTION BY INTRINSIC
SAFETY "I" (IEC 60079-11:2006 (EQV) +
CORRIGENDUM DEC. 2006 (EQV))

Special Conditions for Safe Use:

Misuse

Should the BetaProbe TI/TI+ be overheated or exposed to sudden
physical shock (i.e. being dropped) it should be examined for any
damage that may cause a safety concern. If in doubt please return the
unit for evaluation to Martel Electronics. Please refer to the Customer
Service Section for contact information.

WARNING

Check for proper operation of the instrument before taking it into a
hazardous area. Should the instrument not operate as expected do
not take the instrument into a hazardous area and return it to the
manufacturer for inspection.

WARNING

The BetaProbe TI/TI+ is intended for installation only in locations
providing adequate protection against the entry of solid foreign objects
or water capable of impairing safety.

Caution

To avoid possible damage to calibrator or to equipment

under test:

If the message changes to “OL” the range limit is exceeded and the
BetaProbe TI/TI+ must immediately be removed from the heat source to
prevent damage.

Symbols Used
The following table lists the International Electrical Symbols. Some or all
of these symbols may be used on the instrument or in this manual.

Symbol Description

Power OFF

Power ON

Earth ground

Risk of Danger. Important information. Refer to manual.

Battery

Hazardous Voltage

Conforms to ATEX requirements

Conforms to relevant European Union directives.

Wheeled bin, conforms to EC directive 2002/96/EC

Hazard Location Information/Approvals

Ex Hazardous Areas

An Ex-hazardous area as used in this manual refers to an area made
hazardous by the potential presence of flammable or explosive vapors.
These areas are also referred to as hazardous locations.

Ex ib IIB T4 Gb (–10°C ≤ Ta ≤+50°C)

ITS10ATEX27114X

Ex ib IIB T4 Gb

II 2 G IECEx ITS10.0049X

0344

Manufactured by Martel Electronics, Inc.,

3 Corporate Park Dr.

Derry, NH, USA

2

3

Advertising