WARNING

The RS-232 interface must not be used in hazardous areas.

5. Cleaning

To clean the BetaProbe TI/TI+ use a cloth with a mild cleaning solution.

6. Calibration

The BetaProbe TI/TI+ can be calibrated in the field by one of three
methods depending on the extent of the calibration needed.
1) Single point Ro calibration (typically 0C)
2) Two-Point calibration Ro and gain (typically 100C).
3) A full 4 point calibration.

For complete calibration instructions please contact Martel Electronics
Corp.

www.martelcalibrators.com

4. Changing the Batteries

WARNING

Explosion hazard

Batteries must only be changed in an area known

to be non-hazardous.

Remove the battery door to expose the three AAA batteries. Replace
only with AAA batteries approved in this document. Replace the battery
door and reinstall the hex screw. Be sure to use only the approved list of
batteries. Any substitution will void the safety ratings of this product.

Battery

Manufacturer

Type

Alkaline AAA

Duracell

LR03/MN2400

Alkaline AAA

Rayovac

LR03/824

Alkaline AAA

Energizer

LR03/E92

Alkaline AAA

Panasonic

LR03X

Battery life
Battery life is about 300 hours (12.5 days) of continuous use with the
backlight off. There is a low battery icon in the lower right of the display.
It will appear when battery level is low. Replace batteries per
recommendations found in the specifications section of this manual.
RS-232 Interface
An RS-232 interface is standard on the BetaProbe TI/TI+. Serial
communication can be used for configuration, calibration, and to transfer
measurement data from the probe. When the data logging software is
purchased, a RS-232 and USB cable is included. For customers wishing
to do calibration only the RS-232 cable is available from Martel. Order
P/N 1919069.