Warning – Seametrics MJ-Series User Manual

Page 2

Advertising
background image

Position.

MJ-Series meters should be installed horizontally

with the register up. Vertical mounting will result in some de-

gree of under-measurement and shortened life of the bearings.

No upstream straight pipe is required.

Couplings.

Male NPT threaded couplings are included with

each meter. The threads on the end of the meter are IPS

straight threads one size bigger than the meter size. Though

it is possible to thread a standard pipe coupling directly onto

the meter for close coupling, the included couplings are much

preferable because they provide a union connection for meter

service. Be sure to use the included gasket between the end

of the meter and the coupling.

Connections.

MJE/MJHE and MJR/MJHR sensors are sup-

plied with a color coded output cable (see diagram, page 4).

Optional connectors can be ordered to plug directly into a

Seametrics control or LMI metering pump.

WARNING

DO NOT INSTALL

METER IN OVERHEAD

INDOOR PIPING OR

WHERE LEAKAGE

MAY CAUSE DAMAGE

GENERAL INFORMATION, INSTALLATION and MAINTENANCE

Page 2

GENERAL INFORMATION

MJ-Series meters use the internationally-accepted multi-jet

principle. A gear train drives the register totalizer dials. For

pulse output, one of the pointers is replaced by a magnet

arm, which is detected by an encapsulated sensor attached

to the outside of the lens.

MJE

(cold water) and MJHE (hot water) meters use a solid-

state, long-lasting Hall-effect sensor, which requires power.

They are suited for use with Seametrics controls and metering

pumps (LMI for instance) that have sensor power.

MJR

(cold water) and MJHR (hot water) meters use a reed

switch. They provide a dry contact closure and do not require

power.

MJT

(cold water) and MJHT (hot water) meters do not have a

sensor, and they totalize only.

SPECIFICATIONS*

6 mA at 12 Vdc (MJE/MJHE only)

105˚ F (40˚ C) max
194˚ F (90˚ C) max
150 psi operating

Cast bronze, epoxy powder coated inside and out

Engineered thermoplastic

Alnico

+/- 1.5% of reading

MJE/MJHE

MJR/MJHR MJT/MJHT

Hall-effect

Reed switch Totalizer only

20 mA

20 mA

n/a

24 Vdc

24 Vdc or Vac

n/a

12’ (4 m) standard (2000’ maximum run)

3/4”

1”

1-1/2”

2”

0.22

0.44

0.88

1.98

22

52

88

132

Power

Temperature

Cold Water

Hot Water

Pressure

Materials Body

Internals

Magnet

Accuracy

Pulse Output
Sensor

Max Current

Max Voltage

Cable Length

Flow Rates (GPM)
Minimum
**Maximum

INSTALLATION

Pulse Output.

Both MJE/MJHE and MJR/MJHR sensors

respond to a magnet that rotates on the face of the meter

under the lens. The sensor turns on and off once each time

the magnet passes under it. Sensors are designed for elec-

tronic control loads, and should not be used to switch power

loads or line voltages. See maximum current and voltage

ratings, under Specifications.

MAINTENANCE

Seametrics recommends all service to be performed by an

authorized distributor or the factory to maintain the integrity

of the protective tamper-proof wire-and-seal.

Inlet Strainer.

Clean the strainer yearly, or as required, de-

pending on water condition. Pull out the strainer or backflush

the meter to loosen trapped particulates.

Calibration.

Meters used for billing or billing exemp-

tion may be regulated by state or local authorities. New

meters are factory-tested to meet the AWWA C-708

Multi-Jet Meter accuracy specification. Some states

require retesting at various intervals, typically eight years for

3/4" meters, six for 1", and four for 1-1/2" and 2". Meters

used for control should be tested every 5-10 years. Testing

may be done by a local mobile meter service or in a private

or municipal meter shop. Changes in calibration should be

made at an authorized meter shop.

**CAUTION:

Excessive flow can cause breakage.

Do not exceed recommended maximums.

Caution:

These water meters are

not recommended for installation

in uninsulated suspended ceil-

ings where freezing is possible,

or in any overhead indoor piping

configuration where leakage may

cause damage.

*Specifications subject to change • Please consult our website for current