Seametrics FT400-Series User Manual

Page 10

Advertising
background image

Page 8

SETTINGS

K-FACTOR
At a minimum, every FT400-Series flow computer must
be programmed with the “K-factor”. (This is the number
of pulses that the meter produces per gallon of flow.) If
you wish the FT400 to read in units other than gallons,
see below.

The K-factor on any Seametrics flow sensor fitting or in-line
meter can be found on the model-serial label. The line read-
ing K = xxxx gives the desired number. For depth-adjustable
sensors (110, 210, 150, 250 models), use the calculator
on our website.

Set K.

Begin by pressing the SET key once. The prompt SET K

should appear on the display. The digit to the far right will be blink-

ing. Use the up arrow key to reach your desired value. Then press

the left arrow key to move to the next digit. Repeat the process

until the entire number is entered. (Note that the decimal is fixed

at three places. If you only have two decimal places for your K-

factor, enter a zero for the third digit.) Press SET to advance. (Note:

If unable to set K-factor, the unit is "locked" to prevent tampering.

Please contact your Distributor for assistance.)

Set P/Flow Alarm.

At this screen you may select between pulse

output (P) or flow alarm (A) functions. If the pulse output and

flow alarm features are not being used, this step can be skipped.

The P (pulse output) setting does not affect anything if it is not

being used.

Set P is the default that appears on a new FT400-Series. On an

FT400 that has been previously set up with flow alarm function,

an A will appear on this screen. To move between P and A screens,

firmly press all three keys for 5-10 seconds, then use the up arrow

to scroll through the three options: P, AL HI (high flow alarm) and

AL LO (low flow alarm).

Set 20 mA (FT420 Only).

Press the SET key to advance to SET

20, to set the flow rate, in volume units per time unit, at which 20

mA is desired. Use the up arrow key to reach your desired value.

Then press the left arrow key to move to the next digit. Repeat

the process until the entire number is entered. The processor

will automatically scale the 4-20 mA loop accordingly, with 4 mA

at zero flow.

Set Decimal Point.

Press the SET key again for the D prompt.

Pressing the up arrow key switches among no decimal place, one

decimal place and two decimal places.

Set Time Unit.

When the SET key is pressed again, a

blinking time unit appears. Press the up arrow key

to select SEC (seconds), MIN (minutes), HR (hours) or

DAY (days) (for example, gal/min, or gal/hr).

To save settings and return to normal operation after entering

settings, press SET again. When the unit is connected to an

operating flow sensor, the rate (larger digits) and total (smaller

digits) indicator numbers should appear in the display.

Find Your K-Factor Here

10031295

MF81T-P200

K: 53.6

READING IN OTHER UNITS
Changing Volume Units.

The default K-factor units are

pulses per gallon. To read your total in metric or other
units instead, the standard K-factor must be converted to
the desired volume units. For example, to read in pulses
per liter, the K-factor must be multiplied by the applicable
number shown below.

NOTE: Both rate & total will read in whatever units you choose.

To Convert K to:

Multiply by:

Liters

.26418

Cubic Meters

264.18

Fluid Ounces

.0078

Cubic Feet

7.48

800-Series

Fixed Depth Meter

Changing Time Units:

To read your rate in liters per sec-

ond (for example), convert the K-factor volume units as
shown above and change the time units to Seconds, us-
ing the Set Time Unit instructions at right.

Set P.

From this screen, follow the same process as for Set K

to enter the desired pulse rate. This is the number of gallons

(or whatever units are programmed) between pulses. (Note:

Using the pulse output function disables the high and low flow

alarm functions.) (6.1 Hz max output)

Set

Flow Alarm. From the A screen, use the up arrow key to

choose either AL HI or AL LO and then press the SET key to set

the alarm rate. Use the up arrow and left arrow as above to

reach the desired digits. (Note: Using the flow alarm function

disables the pulse output function.)

Advertising