Seametrics FT400-Series User Manual

Page 9

Advertising
background image

(QUICK) SETTINGS

Page 7

START UP DISPLAY

SET

00001.000

SET K

0.00

MIN

209.8

RESET

0000010.0

SET P

209.8

MIN

PRESS

DISPLAY

SET

SET

SET

K is the number of pulses the flow sensor provides
for every gallon of flow. Find it on the fitting
(800-Series) or chart (100/200-Series) or on the
Seametrics website.

P is the number of gallons per pulse desired on the
scalable pulse output. (Example: P=1 is one pulse
per gallon.) Skip without changing if you are not using
the pulse output.

20 is the 20 mA maximum analog output. Set the
flow rate you want to match maximum output.
Example: 250 gpm maximum expected flow, “set 20”
to 250. The analog output will scale to 4 mA at zero
flow, 20 mA at 250 gpm*.

This setting appears on the

FT420 ONLY.

d is the decimal point. It toggles back and forth with
the . Set as many decimal places as needed. For
higher flows, no decimal allows maximum number of
whole digits.

MIN is the time base, for example, gallons per
minute

. Use the to select sec/min/hour/day.

0001000.0

SET 20

SET

d

Large digits display instantaneous flow rate (GPM).
Small digits display total flow (since last reset).

QUICK SETTINGS OVERVIEW

See following page for step-by-step instructions on changing these settings

Pass through all settings and return to original display to save settings.

*NOTE: Use the up arrow key to reach your desired digit. Then press the left arrow key to move

to the next digit. Repeat the process until the entire number is entered.