Seametrics EM 101 v.2 User Manual

Page 8

Advertising
background image

LT-65200131-C

2/18/10

Seametrics Incorporated • 19026 72nd Avenue South • Kent, Washington 98032 • USA

(P) 253.872.0284 • (F) 253.872.0285 • 1.800.975.8153 • www.seametrics.com

TROUbLEShOOTING

Problem

Probable Causes

Try...

No Display

Flow rate always reads “0”

Flow rate incorrect

Check for 110 Vac
Test fuse, replace if bad
Check ribbon cable connections

Reverse flow connections
Install ground wire
Install meter in vertical position
Use next smaller flow meter
Check connections on display board
Select another flow meter

Check for proper grounding
Select another flow meter
Install meter in vertical position

No power
Blown fuse
Loose ribbon cable

Reversed flow direction
Missing ground wire
Empty pipe
Flow rate below minimum
Loose wiring or incorrect wiring
Fluid conductivity <20 microSiemens/cm

Missing or incorrect ground wire
Fluid conductivity <20 microSiemens/cm
Empty pipe