OPERATION & MAINTENANCE

Page 10

Calibration (“K-Factor”). In order to properly process

pulses from the flow sensor, a number must be entered

into the control to which the sensor is connected. This

number, called the K-factor, is the number of pulses the

sensor puts out per unit of fluid passing through the pipe.

It is normally provided for SeaMetrics sensors in pulses

per gallon, and is given on the chart “K-factors for Various

Pipe Sizes.” These numbers are based on extensive test-

ing, which has shown close agreement between different

EX sensors in the same installation. Typically, most K-fac-

tor error can be attributed to installation variables, such

as depth setting and fitting configuration.

When the EX100 Series Meter is powered up and there is

no flow, there should be no output pulses (or, if connected

to the FT420, flow rate should read “0”). If there are

pulses...

Zero Adjustment. In some cases it may be necessary

to adjust the flow meter under no-flow conditions after it

has been installed.

To per form the adjustment, after determining that there is

no flow, shor t between the two pins marked “Zero Adj.” A

red LED light will come on for approximately 50 seconds

and then go out. The zero adjustment is completed.

Minimum Flow.

As with any other flow sensor, there is a

rate below which the EX100 Series sensor cannot read.

Check the the table on page 11 for the minimum flow rate

detectable by the sensor (0.2) for a given pipe size.

Filtering.

The software of the EX100 Series sensor filters

out electrical noise and also averages sudden variations

in the flow to smooth the output. It takes a matter of

seconds for the flow sensor to get up to full output when it

is powered up or when flow begins. If the unit responds

ver y slowly, it is probably due to excessive electrical noise.

If this is the case, check for adequate grounding to improve

the response.

Electrode Coating.

Grease or other adhering, non-

conductive materials can stop flow detection if the

electrodes become heavily coated. To clean the electrodes,

remove the sensor from the pipe and gently scrub the

electrodes (three dark grey bumps) on the reading face of

the flow sensor. A mild soap (dishwashing liquid for

example) can be used to aid the cleaning process.