Thermo Fisher Scientific Thermo Scientific Temp 16 User Manual

Page 20

Advertising
background image

16

15. CLEAR

FIELD

CALIBRATION


To clear user calibration:
1. Turn the thermometer off.
2. Hold the cal key down while pressing

the on/off key.

3. The measurement mode window

appears without the CAL indicator to
indicate that the user calibration has
been successfully cleared.

16. FIELD CALIBRATION UNLOCK


The calibration unlock feature enables field
calibration operation. To unlock field
calibration:
1. Turn the thermometer off.
2. Hold

the

cal

key and hold key down

simultaneously, then press the on/off
key.

3. The “CAL” indicator appears

momentarily with the version-model
window to indicate that the user-
calibration function has been unlocked
successfully (Figure 10).

CAL

Rtd

Figure 10: Successful field calibration unlock