4
3. SPECIFICATIONS
Platinum RTD Probes
100 ohm platinum: alpha = 0.003850
–200 °C to 850 °C (–392 °F to 1562 °F)
Out of range display: “Ur”, “Or”, “OPEN”
Range Accuracy
–200.0 to –100.0 °C
(–392 to –148 °F)
±2.0 °C
(±4.0 °F)
–99.9 to 199.9 °C
(–148 to 392 °F)
±0.2 °C
(±0.4 °F)
200.0 to 850.0 °C
(392 to 1562 °F)
±2.0 °C
(±4.0 °F)
Resolution
0.1° (–200.0 to 850.0 °C);
0.1° (–392.0 to 999.9 °F);
1° (1000 to 1562 °F)
Display
4-digit LCD (¼ X ½" digits);
2
1
/
3
X 1 ½" backlit viewing
area
Display Update
Rate
0.5 sec per update.
Input
One probe with 3 pin
circular connector
(Switchcraft TA3F)
Input Protection
10V rms
Battery
Three AA, 1.5V alkaline
Battery Life
700 hours continuous
Auto Shutoff
17 minutes after last
keypress.
Stated Accuracy
18 to 28°C
(64 to 82°F)
Useful Range
0 to 40°C
(32 to 104°F)
Storage
-40 to 65°C
(-40 to 149°F)
Humidity
10% to 90%
(non-condensing)
Dimensions
Without armour
With armour
(L x W x H )
175 mm x 97 mm x 42 mm
180 mm x 102 mm x 52 mm