Standardize – Thermo Fisher Scientific CyberScan DO 1500 User Manual

Page 42

Advertising
background image

Instruction Manual

CyberScan DO 1500

39

5.5.

mg/L– One Point calibration (Manual)

User can do calibration in mg/L.


1. Press

std to enter to ‘Standardize’ screen. Both upper and lower

display shows the present measured value.

2. Use

the / keys to adjust the known standard value.


3. Window provided for the adjustment is 70% of the present

reading. Lowest value that can be set is 2.00mg/L; highest value
is 60.00 mg/L.

4. After set the value, press enter key to confirm calibration. Meter

blinks the cal values for few seconds before exit to measurement
mode.

Standardize

press

to

accept

enter

press

to set value

STABLE

mg/L