Thermo Fisher Scientific Ion Selective Electrodes Fluoride User Manual

Page 13

Advertising
background image

Instruction Manual

Fluoride Electrode

- 10 -

where:

x
Cs = concentration of sample
o

Cs = fluoride standard concentration (0.1M)

x
Vt = volume of titrant added to achieve the endpoint of unknown sample

Vt = volume of titrant added to achieve the endpoint in standardization

x

Vf = volume of sample used in sample titration

Vf = volume of standard used in standardization titration.


A typical titration curve is shown in Figure 2.



ELECTRODE CHARACTERISTICS

Reproducibility


Electrode measurements reproducible to

±2% can be obtained if the electrode is calibrated every

hour. Factors such as temperature fluctuations, drift, noise, and variations in illumination limit
reproducibility. Reproducibility is independent of concentration within the electrode's operating
range.

Interference


The hydroxide ion, OH

-1

, is an electrode interference. Anions which make the sample more basic,

such as CO

3

-2

or PO

4

-3

would increase the OH

-1

interference, but do not interfere with direct

electrode operation. Other anions commonly associated with fluoride, such as Cl

-1

, Br

-1

, I

-1

, SO

4

-2

,

HCO

3

-1

, NO

3

-1

, and acetate, do not interfere with correct electrode operation. Most cations do not

interfere with the response of the fluoride electrode to fluoride ion.

Advertising