Instruction Manual

Fluoride Electrode

- 18 -

to complexing agents

Complexation

air bubble on membrane

remove bubble by re-dipping

probe


"Incorrect Answer" incorrect scaling

plot millivolts on the linear axis.

(but calibration

of semi-log paper

On the log axis, be sure that

curve is good)

concentration numbers within

each decade are increasing with

increasing concentration


incorrect sign

be sure to note sign of

millivolt number correctly


incorrect standards

prepare fresh standards


wrong units used

apply correct conversion factor:

10

-3

M = 19 ppm as F

-1


complexing agents

check section entitled

in sample

Complexation