6.5

Temperature Coefficient and Normalization Temperature
Setting

Since different process liquid may require different temperature coefficient factor for
its temperature compensation calculation, COND 500 transmitter allows 0 to 10%
temperature coefficient factor adjustment to cater for your different application
needs. You also have the selection of 25.0 °C or 20.0°C for the conductivity
measurement normalization temperature.

6.5.1

Setting the temperature coefficient

From the measurement mode,

1. Press the ENT key to enter the setup mode. Use

the

▲ or ▼ keys to scroll till LCD displays ‘SEt’

in the upper display; and ‘tPr’ in the lower
display.

2. Press the ENT key to view the temperature

coefficient setup page, ‘tC.O’.

3. Press the ENT key to enter the setup page and

use the

▲ or ▼ keys to set the temperature

coefficient value.

4. Press the ENT key to confirm your setting. The

transmitter will now switch to the normalization
setup page ‘tn.r’

6.5.2

Setting the normalization temperature

From Step 4 above,

1. Press the ENT key to enter the normalization

setup page.

2. Use

the

▲ or ▼ keys to select the desired

normalization temperature (25.0 °C or 20.0°C).

3. Press the ENT key to set the normalization

temperature.

4.

Press

S and T keys together, to return to

the measurement mode.