25

6.10 Reset Function

The reset function lets you choose to either reset the
transmitter’s conductivity calibration only or conductivity
calibration plus all setup functions that you might have
changed back to factory default settings. However, the
temperature compensation setting, as in Section 6.2, will
remain unchanged.

From the measurement mode,

1. Press the ENT key to enter the setup mode. Use

the

▲ or ▼ keys to scroll till LCD displays ‘rSt’ in

the upper display.

2. Press the ENT key to enter the reset page. Use

the

▲ or ▼ keys to toggle between ‘CAL’ and

‘FCT’ for reset type selection.

CAL – reset conductivity calibration only

• FCT – reset conductivity calibration and all

setup functions back to factory default

3. Press the ENT key to confirm your selection.

The lower display will blink momentarily before returning to the reset
function setup mode.

4. Press

S and T keys together, to return to the measurement mode.