Buffer selection settings – Thermo Fisher Scientific Alpha pH 560 User Manual

Page 36

Advertising
background image

Thermo Scientific Alpha pH 560

31

parameters.

Press ENT key to confirm your selection.

If Controller type was set to Off (OFF)

The pH Controller reverts to Cntr screen.

If Controller type was set to On (L.Ct)

3

Select the relay status under Non-Alarm condition:
The lower display shows ‘rEL’. The upper display shows
the last configured relay status. (‘dEEn’ or ‘En’). Press
or key to select the desired relay status.

EEn: De-energized

En: Energized

Press ENT key to confirm your selection. The pH
Controller reverts to Cntr screen.


Press or key to access other setup screens or Press and
key simultaneously (escape) to return to measurement mode.

NOTE: The buffer selection setting is only available if the Controller is
configured for pH measurement mode.

This allows you to set the pH Controller to accept either USA or NIST
pH buffer standard group during pH calibration.

Buffer Group

pH Buffer Points

USA

pH 4.01, 1.68, 7.00, 10.01 & 12.45

NIST

pH 4.01, 1.68, 6.86, 9.18 & 12.45

Figure 5–25. Buffer Selection Settings

Buffer Selection

Settings