Troubleshooting, Chapter 8 – Thermo Fisher Scientific Alpha pH 560 User Manual

Page 45

Advertising
background image

Thermo Scientific Alpha pH 560

40

Chapter 8

Troubleshooting

Problem

Cause

Solution

Power on, but
no display

a) Loose connections
b) Incorrect output

voltage of the
power adaptor

a) Ensure cables make

good contact

b) Use an power

adaptor with
specified output
voltage

Unstable pH
reading

a) Dirty electrode
b) Electrical noise

interference

a) Clean electrode and

recalibrate

b) Switch to Symmetric

mode

Oscillating
temperature
readings

a) Electrical noise

interference


a) Ensure shield wire is

properly connected
to pin 7

Slow response

a) Dirty / Oily
electrode

a) Clean electrode

Blinking ATC

a) No temperature

probe connection
during ATC mode

a) Ensure temperature

sensing cable makes
good contact

Blinking
electrode
annunciator

a) Error in calibration

a) Ensure calibration

standard solution is
not contaminated.
Ensure electrode is
clean.

Or (pH)

a) pH electrode is not

connected

a) Ensure electrode

makes good contact
with Transmitter

Or
(Temperature)

a) Temperature probe

is not connected
when ATC enabled

a) Ensure electrode

makes good contact
with Transmitter