TROUBLESHOOTING

If the software application times out or is unable to update the endpoint firmware for any reason, try the following and then

retry updating the endpoint

IR update errors

Clean the IR head and the endpoint infrared ports with a soft dry cloth to remove any debris or moisture

Disconnect and reconnect the IR programming cable securely to the laptop serial port

When updating an endpoint, block as much of the surrounding light as possible

Try another IR programming cable if one is available

Make sure the IR cable is in good condition Replace it if necessary

Use the optional IR Programming Bracket (PN: 65917-001) to facilitate an IR update

Refer to

"IR Programming Bracket"

on page 46

for details

RF update errors

Check the endpoint

- Make sure the endpoint serial number is valid/accurate
- Make sure the endpoint radio is not in Pause or Stop mode

Multiple updates

- If multiple firmware updates are being performed at the same time, make sure each list of endpoints is

unique—an endpoint should not appear on more than one list

Timed-out error

- If an update times out after several attempts, move the vehicle and retry the update

Check the equipment

Use the switch to turn the mobile transceiver off for a few seconds Then turn it back on

Unplug the mobile transceiver communication cable for a few seconds Then plug it back in

Reposition your vehicle, moving to a better vantage point for the endpoint

Check the software

Check the Settings to make sure the COM ports are valid for the equipment connected to the laptop Refer to

"Settings" on page 13

Close the software application and start it again

If you are still experiencing problems updating an endpoint, contact Badger Meter Technical Support

Technical Support

If you have questions or need assistance with the ORION SE/ME Firmware Updater, contact Badger Meter

Technical Support

Phone: 800-456-5023
Email: TechSupport@BadgerMeter com
Fax: 888-371-5982

TROUBLESHOOTING

Page 47

April 2014

ORION®

SE/ME Firmware Updater