TROUBLESHOOTING

Display Message

Description

Recommended Action

TOKEN NOT

SUPPORTED

Token is not supported.

Token is previously formatted for an alternate feature (for

instance, data logging). New token is required.

TOKEN NOT INSERTED Token operation cannot be performed

because no memory token is detected.

Please ensure the memory token is seated firmly into the

socket located on the amplifier.

TOKEN IS PROTECTED Token is write protected.

Re-programming the token is not

allowed.

Token must be replaced or re-initialized. Re-initializing

token requires special access over MODBUS interface.

Contact tech support for details.

TOKEN IS CORRUPTED Bad checksum found within the

memory token.

Token is not usable in current state. Operator must follow

steps to re-program the token.

Symptom

Recommended Action

Meter is not being restored with contents

on token during power on load.

If meter is secured, un-secure the meter, then retry the restore operation. The

restore operation will re-enable security only if the security pins are stored to

the token. Otherwise, operator must manually re-enable security.

Meter is not being restored with

contents.

Ensure meter firmware is at firmware revision v1.11 or later.

Token Errors are reported in the Error List. Parameter stored in token is corrupted. Token must be re-programmed.

Parameter on token does not exist in meter. Can be ignored if programming a

meter with older firmware.