Badger Meter 4020 Flow Computers User Manual

Page 2

Advertising
background image

©2007- Cox Instruments 830-XX 901TC2

COX 4010 and 4020 Flow Computers

Unit Specifications

Overall Accuracy

± 0.05%

Operating Temperature: 32˚F - 131˚F
(0 - 55˚C)

Environmental

Storage Temperature: -40˚F - 158˚F
(-40˚ - 70˚C)

Power

Nominal Line Voltage: 100, 110, 220
or 240 VAC (50/60 Hz)

Power Consumption

5 Watts max

Temperature Inputs

Compatible RTD type: 100

Platinum

(4020) Only

Lead Wire Compensation: 4 Wire
Configuration 2, 3 or 4 Wire

Range: 0-24 VDC, sink only.

Flow Inputs

Thresholds: High = 2.8 - 24 VDC

Low = 0 - 1.3 VDC

Impedance: 5.8 k

Six Digit Display

Flow Rate Display

3 Character units of measure identifier
(GPM, LPM, PPH, CCM, etc.)

Type: Dry Contact, Form C

Relay Outputs

Contact Rating:
10A@115/230VAC/28VDC

Type: NPN

Transistor Outputs

150 mA Maximum
30 VDC

Auxiliary Power

(AC powered units only)

Outputs

Voltage: 24 VDC regulated and filtered
Current: 0 to 100mA

Analog Output

4-20 mA, 12-27 VDC
ACC: ± .1% F.S. update = 2Hz

Frequency Range

0-7.5 kHz

Communications

RS-485 Multidrop

FLOW COMPUTER APPLICATION

• Turbine flowmeter sends digital signal to

Flow Computer

• 100W RTD direct hook-up to Flow Computer

(4020 Only)

• Flow Computer Linearizes, Temperature

Compensate (4020 Only) and calculates
corrected Flow Rate & Total

• Scaled pulse out to remote totalizer in

supervisory area

• 4-20 mA out, corrected flow rate, to strip chart

recorder tracks trends

• Alarm relay and transition outputs to activate

bell and/or light as needed

For more information, contact COX Instruments

or your local COX Instruments representative.

15555 North 79th Place • Scottsdale, AZ 85260

tel: (480) 922-7446 • fax (480) 948-3610

www.cox-instruments.com

Advertising
This manual is related to the following products: