Blue-White FLEX-PRO A2 ProSeries Peristaltic Pump User Manual
Page 21

Page 21
Flex-Pro
Step 7
To exit Setup, press and hold SELECT RUN MODE button
for a few seconds until you return to Run screen.
Step 6
mA value linked to minimum pump speed will now be
displayed.
To increase value, press and release UP arrow.
To decrease value, press and release DOWN arrow.
To save value, press and release RIGHT arrow.
20
M
A
2
Mode 2
Mode 2
OFF
mA
2
Example 1
6
4
8 10 12 14 16 18 20
0
25
50
75
100
Pump
Motor Speed (%)
Milliamp input (mA)
from external source
4 mA = 0% Pump Output
20 mA = 100% Pump Output
Example 2
6
4
8 10 12 14 16 18 20
0
25
50
75
100
4 mA = 100% Pump Output
20 mA = 0% Pump Output
Example 3
6
4
8 10 12 14 16 18 20
0
25
50
75
100
4 mA = 20% Pump Output
16 mA = 43.8% Pump Output
0%
Output
Pump
Motor Speed (%)
Pump
Motor Speed (%)
100%
Output
100%
Output
0%
Output
43.8%
Output
20%
Output
Milliamp input (mA)
from external source
Milliamp input (mA)
from external source
Runtime Screen Shot 1
Mode 2
9.1
Mode 2 - 4-20mA Input Screen Shots
Runtime Screen Shot 2
Mode 2
9.6
mA
2
9.0
Mode 2 - 4-20mA Input
- Continued
Operation
Display motor speed percentage.
Pump Running in 4-20mA Input Operation
Display current 4-20mA input signal
Press and release RIGHT arrow to view mA input value
in real-time.
35.0
SPEED
2
Runtime Screen Shot 3
Mode 2
9.6
mA
2
Press and release RIGHT arrow again to view mA output
value in real-time
(available on select models only)
.
Press and release RIGHT arrow again to view motor
speed percentage, as in Screen Shot 1.
OUT