Password protection, Loading and saving configurations, Reset to factory defaults – Crystal ConfigXP Digital Pressure Gauge Software User Manual

Page 12: Loading and saving confi gurations

Advertising
background image

Settings

12

Confi gXP User’s Manual

PASSWORD PROTECTION

The password protection feature of the gauge allows you to prevent unauthorized changes to the gauge confi guration by setting a four digit password. Once

the password is set you will not be able to make any changes to the gauge through Confi gXP without the password. In addition, any functions within the

gauge itself that can be accessed by the keypad or serial interface, such as water reference select or userspan adjustment, will be disabled. Any serial interface

command that attempts to change a protected feature will return a

X,0

message indicating that feature cannot be changed. Passwords can only be set or

removed using Confi gXP.

In the event you lose the password, you will need to contact the factory for an unlock code, which will remove the password protection.

LOADING AND SAVING CONFIGURATIONS

Use the

Load Confi guration

and

Save Confi guration

buttons to create custom setups–like enabling a limited set of engineering units or enabling other

features in the gauge–and storing those confi gurations for later use. Applications might include using a saved confi guration to setup multiple gauges identi-

cally, or using diff erent confi gurations for the same gauge to customize it for diff erent applications. Confi gurations do not include the password or userspan

value.

RESET TO FACTORY DEFAULTS

The

Load Factory Defaults

button will return the gauge to the confi guration as shipped from the factory. As part of the reset function, the password will be

removed from the gauge, but the userspan value will be unaff ected.