Units button options – Crystal ConfigXP Digital Pressure Gauge Software User Manual

Page 8

Advertising
background image

Settings

8

Confi gXP User’s Manual

UNITS BUTTON OPTIONS

Select the pressure units you need, and turn off the pressure units you never use (at least one unit must be selected).

Use the Unit button options to control which units the pressure gauge will display.

Default Unit:

The pressure unit displayed after a reset, and the unit used when in datalogging mode. (Reset occurs when the batteries are changed, when an

!RST command is sent to the gauge over the serial port or, on older units, when the calibration settings are changed.)

Water Density:

From the dropdown menu, select

4 °C (39.2 °F)

,

15.6 °C (60 °F)

, or

20 °C (68 °F)

. Aff ects inches of water (inH

2

O), only.

User Defi ned Units:

This section allows you to create a custom pressure unit. When you select this unit the gauge will display

USER

. The display will equal

the User Factor multiplied by the measured psi plus the User Off set. Use the

Resolution

dropdown menu to select the number of decimal places on the

display. Confi gXP automatically limits the maximum number of decimal places depending on the pressure range of the gauge, so that only valid numbers

can be displayed on the gauge.