Overview, Introduction, Capabilities – Crystal ConfigXP Digital Pressure Gauge Software User Manual

Page 3

Advertising
background image

Overview

3

Confi gXP User’s Manual

Overview

INTRODUCTION

Confi gXP provides an easy way to customize your gauge (DTG, XP2, or XP2i). With only a serial cable (P/N: 2400, or equivalent*), a computer running Windows,

and the Confi gXP program you can:

View and print an

as receive versus as left change report

Load and save custom confi gurations

Return the gauge to factory confi guration

Select pressure units (e.g., disable inches Hg, if you never use this scale)

Adjust (calibrate) the gauge

Identify fi rmware version

Note:

DTGs require an RS-232 interface adapter kit (P/N: 2746).

CAPABILITIES

Some features are dependant on the model and fi rmware version of your device. You may update your fi rmware using

XP2iUpdate

.

For many gauges, you will be able to:

Password protect the gauge to prevent unauthorized changes

Disable (or enable) the

(

peak

)

button

Select your preferred H

2

O water density: 4°C, 60°F, 68°F/20°C

Defi ne a special pressure unit (e.g., display pressure as torque)

Store a message or identifi cation number

Limit or expand the zero adjustment range

Program a TARE value

Enable and adjust display damping (averaging)

Display rate of pressure change

Disable the ability to adjust (calibrate) the gauge from the keypad

Note:

Features not available on your gauge will be grayed out in Confi gXP.

!

CAUTION: Straight pass-through RS-232 cable required. DO NOT use a null-modem cable.

!

CAUTION: Changing the settings of your gauge using this program can cause the unit to display inaccurate readings or behave diff erently than the

factory defaults. Only personnel qualifi ed to make changes should do so.

Advertising