Page 7
CONNECTIONS DIAGRAMS (WMX101)
The FLMG requires a power source of 8-32 VDC at 30 mA
max (see WARNING). The power cable also serves as a pulse
output if needed for remote reading, data logging, signal
conversion, or telemetry.
Series RTI/ Series FLMG
Orange and Blue
: Serial Output (Do Not Use)
Green (+) and White (-)
: Pulse Output, 30 Vdc max, 10 mA max
Red (+) and Black (-)
: External Power,
8-32 VDC at 30 mA max
Drain
: Connect to earth ground (see WARNING)
Series BAT/ Series FLMG
Cable Color Codes
WARNING:
Using an unregulated power
supply >18 Vdc may damage the meter
due to AC line input voltage fluctuation.
WARNING:
Using an unregulated power
supply >18 Vdc may damage the meter
due to AC line input voltage fluctuation.