EQUALIZATION AND GROUNDING

Metal Pipe Installations.

To equalize the electrical potential

of the fluid, the FLMG meter, and the surrounding pipe, secure
the flange plates (factory-installed on the equalization lug) to
both pipe flanges at one of the bolt holes, as shown below.
Be sure the lockwasher fits between the pipe flange and the
flange plate.

Equalization Diagram

Equalization Lug

Meter

Flange

Pipe

Flange

Bolt

Run wire from equalization lug to both pipe flanges;

secure flange plates under bolt heads as shown.

Meter Flange

Pipe Flange

Flange Plate

Lockwasher

WARNING: ELECTRICAL SHOCK HAZARD

When

the meter is externally AC powered, the piping sys-

tem must be grounded to meet national and local

electrical safety codes. Failure to do so can result

in electrocution.

Plastic Pipe Installations.

When the FLMG is installed in a

plastic piping system, it is not necessary to use the equaliza-
tion straps, but very important to ground the meter to avoid
electrical shock hazard and electrostatic interference with
meter function.

INSTALLATION

Positioning the Meter.

These meters can be installed

horizontally, vertically, and in any radial position.

Straight Pipe Recommendations.

As with most flow

meters, the FLMG requires some straight pipe before and
after the meter for best accuracy. However, the ability of
electromagnetic meters to average the flow across the entire
pipe allows for shorter straight pipe recommendations than
most mechanical meters (see page 4).

Full Pipe Recommendations.

All magmeters require a

method for determining that the pipe is empty, to prevent false
reading. This meter is designed to go to zero reading if one
or more electrodes is exposed. For highest accuracy, install
the meter so that the pipe will be full when there is flow. If air
bubbles may be present in the pipe or sludge accumulation
is an issue, rotate the meter by one flange hole to position
the control housing at a 45˚ angle (see diagrams on page 5).

Fittings.

The FLMG flanges have standard ANSI 150 lb drilling

and mate with any other ANSI 150 lb flange.

Calibration.

The FLMG is factory-calibrated and will not require

any form of field calibration.

Caution:

These flow sensors are not recommended

where installation fault may expose the flow sen-

sor to boiler pressure and temperature. Maximum

recommended temperature is 130˚ F.

INSTALLATION and GROUNDING

Caution:

In chemical injection ap-

plications, install chemical injection

point downstream of magmeter, or far

enough upstream to allow complete

mixing of fluids before the meter.

Chemical Injection.

When any magmeter, by any manufac-

turer, is used in a chemical injection application, the chemical
injection point must be placed downstream of the magme-
ter OR far enough upstream for complete mixing to occur
before the fluid reaches the meter

. When unmixed chemical

alternates with water passing through the meter, the rapid
changes in conductivity may cause sudden spikes and drops
in the meter’s reading, resulting in inaccurate measurement.
The magmeter will restabilize, however, with a steady flow of
fluid of uniform conductivity.