Dwyer FLMG User Manual

Page 11

Advertising
background image

Problem

Probable Cause

Try...

Blank Display

Flow rate steadily reads

zero when there is flow

Flow rate intermittently

drops when there is flow

Jumpy reading

Meter reads, but no

pulse output

Output pulses missing

Replace battery pack, check power

connections

Reading will resume when flow increases

Reposition meter for full pipe (see page 4)

Note flow direction arrow, reverse meter

Change power connections

Select another flow meter

Reposition meter for full pipe (see page 5)

Check for proper ground

Use external power source

(allows more flow averaging)

Install chemical injection line downstream

of meter (or far enough upstream to allow

complete mixing of fluids before meter)

Add pull-up resistor

Change output connections

Check display

No power plus dead backup battery

Flow is below cutoff (very low)

Pipe is not full

Meter is installed backwards

(display reads [ - ] )

Power connections reversed

Fluid conductivity <20microSiemens/cm

Pipe is not full

Missing or incorrect ground wire

Pulsing flow

Rapidly changing conductivity

(in chemical injection applications)

External device needs pull-up resistor

Reversed leads (polarity sensitive)

Meter not reading

TROUBLESHOOTING