______________________________________________________________________________

Kaman Precision Products

PART NO: 860522-001

www.kamansensors.com

Last Revised 11/16/2012

- 12 -

R112 Gain Ch X


R117

Offset

Ch

X

R

217 Offset Ch Y

R212 Gain Ch Y

Figure 9 Potentiometer Locations



Note:

Offset (zero) potentiometers R117 and R217 are not normally installed and are not

included in this calibration sequence.

1. Position the target at Null

2. Adjust the offset control for the measuring channel being calibrated to 0+/-0.001V (unless

otherwise specified).


3. Position the target at –FS (negative full scale)

4. Adjust the gain control for the measuring channel being calibrated to the full scale voltage +/-

0.005mV (unless otherwise specified)


5. Position the target at Null

6. Adjust the offset control for the measuring channel being calibrated to 0+/-0.001V (unless

otherwise specified). (See above note if there is no offset control)


7. Position the target at –FS (negative full scale)

8. Adjust the gain control for the measuring channel being calibrated to the full scale voltage +/-

0.005mV (unless otherwise specified).


9. Repeat steps 1-8 as necessary