8000 SERIES OPERATION MANUAL

TRANSMILLE LTD.

Page 45

2-Wire Resistance (Electrometer Mode) (Option)


Press

Ω 2WIRE to enter the Ω 2WIRE measurement mode – the status display will indicate the

settings and range information.

Selecting ranges above 10 MOhms will activate electrometer mode - this is displayed in terms of
current (100uA, 10uA, 1uA, 100nA, 10nA). Inputs are measured from the IMeas BNC terminal on the
right hand side of the multimeter. An LED indicates the BNC terminal set is active.

Voltage output on electrometer terminals is configurable from status menu (using menu dial) from
50V to 300V in 50V steps.

Status Display Section

Description

X RANGE

eg. *10m

Ω DC

Display / Selection of 2-Wire Resistance range –

Use menu dial or

RANGE UP RANGE DOWN

buttons to change range.

Selected range indicated by a *

MANUAL

Ranging mode : AUTO or MANUAL

FRONT

FRONT or REAR terminals in use

Additional Settings

Description

DIGITS

Set measurement resolution

FILTER TIME

Set FILTER TIME : 250ms to 64s

Dynamic Mode for filter setting based on measurement stability

ANALOGUE FILTER

Set ANALOGUE FILTER mode :

ON / OFF

READING MIN / MAX

Display READING MIN / MAX measurement

TRIGGER

AUTO / MANUAL

GUARD TO LOW

Set GUARD TO LOW mode : FLOATING / SIGNAL LOW

MATHS, NULL, dB

Set NULL : Store zero offset value

Clear NULL : Clear zero offset value

Set OFFSET : Add a user entered value to the measurement

Set SCALE : Multiply measurement by a user entered value

Clear FORMULA : Clear the formula entered (SCALE & OFFSET)

See dB function for use of Set dB, dB SET ZERO & dBm REF 600 Ohms

POWER / CURRENT

Displays the power dissipation / current flowing

for the specific range & measurement

TEST VOLTAGE

Select Test Voltage 50V • 100V • 150V • 200V • 250V • 300V

for the specific range & measurement

INTERNAL

TEMPERATURE

Displays the internal temperature